Fluke 719P-300G/KIT -0.8 bar to 20 bar Pressure Calibrator

小計(1 件)*

TWD218,686.00

(不含稅)

TWD229,620.30

(含稅)

Add to Basket
選擇或輸入數量
有庫存
  • 1 件準備從其他地點送貨
  • 加上 2 件從 2026年12月16日 起發貨
**需要更多嗎?**輸入您需要的數量,然後按一下「查看送貨日期」以查詢更多庫存和送貨詳細資訊。
單位
每單位
1 +TWD218,686.00

* 參考價格

RS庫存編號:
458-124
製造零件編號:
FLK-719P-300G/KIT
製造商:
Fluke
透過選取一個或多個屬性來查找類似產品。
選取全部

品牌

Fluke

Battery Type

AA Alkaline

Maximum Pressure Measurement

20bar

Product Type

Pressure Calibrator

Minimum Pressure Measurement

-0.8bar

Accuracy

0.001 bar

Resolution

0.001 bar

Pressure Port Type

Gauge

Model Number

719P-300G/KIT

Display Type

Backlight

Current Loop Measurement

Yes

Leak Testing

Yes

Power Source

Battery

Minimum Operating Temperature

-10°C

Weight

1.2kg

Maximum Operating Temperature

50°C

Length

70mm

Width

111.8mm

Standards/Approvals

Traceable Certificate of Calibration

Height

241.3mm

Hazardous Area Certification

No

COO (Country of Origin):
US

Fluke Pressure Calibrator, 20 Bar Maximum Pressure Measurement, 0.001 Bar Accuracy - FLK-719P-300G/KIT


This pressure calibrator is a battery-powered handheld instrument designed for on-site calibration and testing of transmitters, switches and gauges within industrial environments. It combines an integrated electric pump for single-handed pressure generation with mA loop sourcing, simulation and measurement, and a large backlit display for simultaneous readouts, allowing technicians to perform pressure, electrical and temperature checks under field conditions.

Features and Benefits:


• 0.001 bar resolution for precise pressure adjustment and verification
• Accuracy 0.001 bar enabling high-precision transmitter calibration
• Programmemable pump limits to prevent over-pressurisation during tests
• Current loop measurement and sourcing for 4-20 mA loop tasks
• Leak testing function to confirm system integrity after interventions
• Backlight display for clear readings in low-light environments

Applications


• Suitable for calibrating pressure transmitters on site
• Used with pressure switches to verify trip settings
• Ideal for validating gauge accuracy in process plants
• Can be used for servicing pneumatic control systems
• Used for loop checks and signal simulation during commissioning

What pressure range can it measure and generate?


It measures down to −0.8 bar and up to 20 bar with selectable pressure ranges including 30 psi, 150 psi and 300 psi equivalents.

How is the unit powered and what is battery performance?


It uses eight AA alkaline batteries and provides a minimum of 1,000 pump cycles to 30 psi and 300 pump cycles to 150 psi depending on range.

Which connections support external sensors and accessories?


Pressure connection is a 1/8 in female NPT and electrical connections use standard banana jacks Plus dedicated RTD and external module sockets.

How does temperature affect measurements?


Add ±0.002% full-scale per °C for conditions outside 15 °C to 35 °C, with no accuracy change within that band.

What is included to protect the pump from contamination?


A fluid trap is supplied and pump cleanout ports are provided to reduce moisture and fluid damage.