Fluke Multimeter Kit, 10A Current, Electrical Maintenance Application - FLUKE 117/PRV240


This multimeter kit is a handheld electrical test set designed for maintenance and diagnostic tasks in industrial and commercial environments. Supplied with test leads and a proving unit, it supports a wide range of voltage, current, resistance and frequency measurements, and is intended for routine electrical maintenance where portable, safety-rated measurement equipment is required.

Features and Benefits:


• 10A capacity protects against high-current tasks
• True-rms measurement for accurate AC readings
• 600V maximum between Terminal and earth ground
• 6,000-count display with 33-segment bar graph updates
• 9V alkaline battery providing long runtime without backlight
• IP42 ingress protection for routine site use

Applications


• Suitable for electrical maintenance on industrial switchgear
• Ideal for diagnostic testing of control panels and circuits
• Used for measuring AC and DC voltages up to 600V
• Can be used for continuity checks and diode testing
• Suitable for capacitance and frequency verification tasks

What safety standards does the kit meet?


It holds approvals including TUV, CSA, VDE, CE and UL and meets IEC 61010-1 with CAT III 600V, 10A ratings.

How does the unit handle temperature extremes during storage and operation?


It operates from -10°C to +50°C and can be stored between -40°C and +60°C.

What input protections and fusing are provided for the current ranges?


The A input uses an 11A, 1000V fuse (where fitted) with high interrupt rating and the design limits overload to 20A for 30 s on designated ranges.

How is measurement accuracy maintained across varying humidity and altitude?


Accuracy is specified for one year after calibration at 18°C-28°C, humidity and altitude operating limits are defined to ensure specified performance.